- Multi parameter patient Monitor.
- Diagnostic quality of ECG.
- Advanced digital filtering.
- Onsite upgradability up to 5&7 parameters.
- Inbuilt rechargeable Battery (Optional).
- Serial Data communication.
- NIBP, ECG-12 lead, pSO2, all medical equipment compatible DICOM can transfer data/images. By this equipment therefore mentioned Data can be transmitted live. Since it is scalable, any DICOM compatible unit, for e.g.: Electronic Stethoscope can be attached to it.
Technical Specifications
ECG
No of leads : 12
Mode : Differential
Input Impedance : > 10 M ohms
Leakage current : < 10 micro amps
Frequency response : 0.05 to 40 Hz (Digital)
SpO2
Range : 0-100 %
Accuracy : + 2% for 70-100%
+ 3% for 50-69%
Pulse rate range : 30 to 254 BPM
Pulse rate accuracy : + 3BPM
NIBP
Range : 10-300 mmHg
Accuracy : + 3 mmHg
Measurement time : 30 to 150 seconds
Measurement Intervals : Programmable
Systolic : 30 - 290 Programmable
Diastolic : 20 - 290 Programmable
General
Dimension : 263L X 180W X 70H mm
Weight : < 1.5 Kgs
Power : 230 V, 50 Hz
Battery : 6V, 3 Ah (NlmH)
Operating time : 2 Hrs
Charging time : 10-12 Hrs.
PC
CPU, Pentium-IV,2.5Ghz and above, 512 MB RAM, 80GB HDD, CD Drive, 15" LG Black Monitor, T.V.Tuner Cord, Optical Mouse, Keyboard, Normal with Hiroshima cabinet. |
|